// Intraoperative Visual Identification of Extra-Cochlear Electrodes in Cochlear Implantation Using Stimulation Current-Induced Non-Stimulating Electrode Voltages
本研究評估刺激電流誘發非刺激電極電壓(SCINSEVs)作為耳蝸植入手術中快速偵測耳蝸外電極(EEs)的方法|結果顯示敏感度達95.9%,特異度為40.7%,證明SCINSEVs能即時發現誤置電極|此技術有助減少術中影像依賴,提升手術精準度與患者預後。
// This study evaluates Stimulation Current-Induced Non-Stimulating Electrode Voltages (SCINSEVs) as a fast intraoperative method to detect extra-cochlear electrodes (EEs) during cochlear implant surgery. Researchers recorded SCINSEVs while deliberately leaving electrodes outside the cochlea and compared expert reviewers’ detections with surgical observations. Results showed high sensitivity (95.9%) but moderate specificity (40.7%) for identifying EEs, suggesting SCINSEVs reliably detect misplaced electrodes in real time. This approach may reduce reliance on imaging and improve outcomes by enabling immediate surgical correction.