Audiology and Neurotology2026年8月25日

機器人輔助電極陣列植入對人工電子耳受者神經健康與聽力結果之影響

// Impact of Robotic-Assisted Electrode Array Insertion on Neural Health and Hearing Outcomes in Cochlear Implant Recipients

本研究比較手動與機器人輔助的耳蝸植入電極置入,涵蓋20名患者|透過神經反應電位(ECAP)與語音辨識測試,機器人輔助法在神經保存與語音表現上均顯著優於手動|結果顯示機器人輔助手術能有效提升植入效果,促進神經結構保護與聽力表現改善。

// This study compared manual versus robotic-assisted cochlear implant electrode insertion in 20 patients to assess effects on neural health and auditory performance. Using ECAP measurements and speech intelligibility tests, robotic-assisted insertion showed significantly better neural preservation and higher speech scores than manual insertion. A positive correlation was found between neural health (ECAP slope) and speech intelligibility outcomes. The results indicate robotic-assisted surgery improves cochlear implant effectiveness by better preserving neural structures.

Published 2026年8月24日